Denoising Effect on T2* Values in Magnetic Resonance Imaging with Application in Iron Load of Patients with Thalassaemia Major


1 Postdoc
2 Radiology Department, Shiraz University of Medical Sciences
*Corresponding author: Mahdi Dodangeh, E-mail: dodangeh@gmail.com.
Iranian Journal of Radiology. 2017 May; Special iss(5): e48196 , DOI: 10.5812/iranjradiol.48196
Article Type: Abstract; Received: Dec 21, 2016; Accepted: Feb 8, 2017; epub: Apr 13, 2017; ppub: May 2017

Abstract


Objectives: To investigate the effect of noise and denoising on the measurement of magnetic resonance (MR) imaging T2* values at interventricular septum of the heart in thalassemia major patients. A second objective is to estimate the T2* value based on the improvement of signal to noise ratio (SNR).

Methods: Different levels of Rician and Gaussian noises were added to the cardiac T2* MR images of 20 thalassemia major patients. The state-of-art denoising methods were applied to the obtained noisy images. T2* values of the interventricular septum and SNR were measured in both series of images by Segment and Matlab software, respectively.

Results: With respect to the Gaussian noisy and denoised images in different noise levels, although SNR improved, there were no significant differences between T2* values in original, noisy and denoised images. On the other hand, after denoising images with Rician noise a relation between improvement of SNR and T2* value is noticed. A predictor model is constructed based on the level of Rician noise for each method.

Conclusions: In the case of existence Gaussian noise, the results confirm that denoising is not effective on the measurement of T2* value. In the case of image distortion by Rician noise, a predictor model is proposed to estimate the original T2* value. The predictor model is used to estimate the T2* value of new patients. The predicted T2* values were in good agreement with the corresponding original T2* values.

This is an abstract presented in the 33rd Iranian congress of radiology (ICR) and the 15th congress of Iranian radiographic science association (IRSA).